qiao@hvtest.cc15871365102
  • Analysis and Remedial Measures for Common Malfunctions of Chromatography Instruments

    The oil chromatography analyzer under UHV power can help many power workers conduct various power tests more conveniently.


    16


    1、 The principle and application of chromatographic analyzer

    The chromatographic analyzer mainly detects the characteristic information of different components contained in natural gas, in order to obtain accurate natural gas component content. In practical applications, the chromatograph will perform multiple adsorption and desorption operations, analyze the time of appearance of the chromatogram, and record the content of components in the backend system to meet the requirements of natural gas component detection.

    2、 Analysis and Remedial Measures for Common Malfunctions of Chromatography Analyzer

    (1) Airway pollution. Chromatography equipment is prone to gas path pollution in daily work. The main reason is that natural gas contains high concentrations of impurity gases. The content of impurity gases can be visually observed through charts. The main reasons for the presence of impurity gases in natural gas are: contamination of the inlet and outlet of the chromatograph detection pipeline, resulting in fixed peaks at both ends of the graph; The appearance of non periodic chaotic peaks in the figure indicates the presence of multiple impurity gases; When the internal flow direction of natural gas changes, it can cause gas path pollution in the chromatograph. Solution: Staff should check the actual location on the map and clean the area; When cleaning, use clean air to blow the inlet and outlet areas of the chromatograph. If the map shows a high content of impurity gases in the area, a mixture of nitrogen and ethanol should be used for operation; When there is a large amount of polluted gas, it is necessary to use the method of heating aging; When the detector is slightly contaminated, the ethanol nitrogen method can be used instead.

    (2) The separation of components is not ideal. If the component separation is not ideal during the actual operation of the chromatograph, the main reason may be the unreasonable column temperature setting. Solution: It is necessary to conduct experiments on the column temperature of the chromatographic analyzer to obtain an accurate dynamic change chart of the column temperature, ensuring the rationality of the column temperature.

    You Might Also Like

    • Have Any Questions!

      DON'T HESITATE TO CONTACT US ANY TIME
      Contact us
    • Visit Us

      Building 6, Optics Valley Optical-Mechanical Industrial Park, 84 Gaoxin 5 Road,
      East Lake New Technology Development Zone, Wuhan City, Hubei Province, China

    • Email Us

      qiao@hvtest.cc

    • Call Us

      +86-15871365102